OBJECTIVE: The aim of this report is to propose a scheme for validation of an analytical technique according to ISO 17025. MATERIAL AND METHODS: According to ISO 17025, the fundamental parameters tested were: selectivity, calibration model, precision, accuracy, uncertainty of measurement, and analytical interference. RESULTS: A protocol has been developed that has been applied successfully to quantify zinc in serum by atomic absorption spectrometry. CONCLUSION: It is demonstrated that our method is selective, linear, accurate, and precise, making it suitable for use in routine diagnostics.


Assuntos
Análise Química do Sangue/normas , Espectrofotometria Atômica/métodos , Zinco/sangue , Acreditação/normas , Calibragem , Humanos , Agências Internacionais/normas , Variações Dependentes do Observador , Reprodutibilidade dos Testes , Sensibilidade e Especificidade , Espectrofotometria Atômica/instrumentação , Espectrofotometria Atômica/normas